FOX6-XM
5V / 1400mA worldwideQuantity | Unit price |
---|---|
To 9 | €23.94 * |
From 10 | €20.06 * |
From 25 | €16.64 * |
- Order number: 1898096
- Availability:
-
Delivery time:
Germany 2 - 3 working days, Europe 4 - 5 working days, worldwide 10 working days
Maximum order quantity: 750 pcs
Efficiency and standby losses in accordance with DoE 2016 IEC 60601-1 approved ≤ 10µA leakage... more
Required Accessories
For the operation of this device you need a primary adapter and a secondary adapter . This can be found here:
Primary adapters and Secondary adapters
Primary adapters and Secondary adapters
Product information "FOX6-XM"
Efficiency and standby losses in accordance with DoE 2016
IEC 60601-1 approved
≤ 10µA leakage current
Protection class 2xMOPP
DOE 2016 ready
IP42 approved exchangeable plug system as additional accessories
Double Safety: Housing snapped & welded
IEC 60601-1 approved
≤ 10µA leakage current
Protection class 2xMOPP
DOE 2016 ready
IP42 approved exchangeable plug system as additional accessories
Double Safety: Housing snapped & welded
Additional information
- Output voltage: 5 V
- Current: 1400 mA
- Gehäuse / Housing: FOX6
- Frequency: 50 / 60
- AC Input: FRIWO interchangeable mains plug system: see Primary Adapters
- Humidity: 10% bis 90%
- Working temperature: 0°C - 45°C
- Storage temperature
: -40°C - 70°C - DC output: Universal output plug system
- Input tolerance: ±10%
- Input voltage: 100-240 V
- Protection Class: II
- Spec.0 cover sheet FW-Type: FW8002M/05
- color: black
- Input current: 160-80 mA
- leakage current: ≤10 µA
- Overload protection: Yes
- Over Voltage Protection: Yes
- Short Ciruit protection: Yes
- Operating Altitude: 3000 m
- Dimension: 55x31,5x41 mm
- material: PC-ABS
- Length: 1830 mm
- Standard: IEC60601-1
- Output voltage tolerance: ±5%
- Weight: 112 g
- Storage Humidity: 10-95%
- Inrush current: <35 A
- Atmospheric pressure: 70-106 kPa
- Line Regulation: ±1
- Load Regulation: ±4
- Hold up Zeit: >5ms(120V) ; >50ms(230V)
- Housing Flame Class: UL 94 V0
- IP Rotection rating: IP40/(IP42 with special Adapter)
- Condcted and Radiated Emmisions: EN55022, EN55011, FCC15, Class B, EN60601-1-2
- Immunity
: EN55024, EN60601-1-2 - Harmonics: EN61000-3-2
- Flicker noise: EN61000-3-3
- ESD (contact/ air): EN61000-4-2
- Immunity against radiated field: EN61000-4-3
- ETF/ Burst: EN61000-4-4
- Surge: EN61000-4-5
- Immunity against conducted disturbances: EN61000-4-5
- Voltage dips: EN61000-4-11
- Immunity against magnetic field: EN61000-4-8
- Drop test: Standard: IEC60068-2-32, Surface: steel, Drop height: 1m, Numbers of Drops (total): 3
- Vibration: Standard: IEC60068-2-6, Frequency range: 10-500Hz, Cross-over-frequenzy:58-62Hz, Displacenebt amplitude:0.2mm, Peak acceleration: 3g, Number of cycles: 10 per axis
- Means of Protection: Input Output 2x MOPP, Input- Ground 1xMOPP
- MTBF: 200.000h acc. Mil217F (based on calculations at
120Vac/60Hz & 230Vac/50Hz, ambient 25°C
and 100% load) (The MTBF are theoretically determined values, which does not guarantee the lifetime of the product or of the electrolytic capacitors) - Approbations: Europe, USA, Australia, Japan
Documents
Service Software
Software License FRIWO Enable Tool Application / 1 User, 1 Year / 1 Project code